Recall Z-2467-2021 ยท FPA Set, Administration, Intravascular

CAREFUSION

BD SmartSite, REF: 2000E CHINA; BD SmartSite Connector, REF: 2000E-04; BD SmartSite J-LOOP Extension Set One Needle Free Valve, REF: 20021E7D; BD SmartSite Extension Set One Needle Free Valve Slip Luer, REF: 20037E7D; BD SmartSite Extension Set 3 Needle Free Valves, REF: 20038E7D; BD SmartSite Extension Set One Needle Free Valve, REF: 20039E7D; BD SmartSite Extension Set 1 Needle-Free Valve, 20049

DateAug 2, 2021
StatusOpen, Classified
ReasonNeedle-free connector valves, either standalone or attached to extension sets, may have issues including difficulty to flush, flow issues, and partial or total occlusions, which could lead to therapy delay. Follow-up activates not required if product has already been used or was safely connected to the female luer of connecting device.
Root causeProcess control
ActionOn 08/02/21, Urgent Medical Device Correction notices were mailed and emailed to consignees who were asked to notify recalling firm if this issue was experienced at 1-844-823-5433, productcomplaints@bd.com . Follow-up activates not required if product has already been used or was safely connected to the female luer of connecting device. If during priming the clinician encounters an occluded needle-free connector, take the following actions: 1) Depress the syringe plunger while the syringe is still attached to the needle-free connector. 2) If unable to depress the plunger, pull back on the syri
Quantity23,786,178
DistributionWorldwide distribution - US Nationwide distribution in the states of FL, PA, NC, OH, SC, IL, RI, MO, NY, TN, AL, VA, PR, CA, NE, NJ, MI, MD, NM, LA, KY, NV, ME, OK, WA, MA, WV, TX, AR, OR, KS, GA, VT, IN, WY, AZ, CO, MN, WI, UT, SD, MS, ID, MT, ND, HI, IA, DC, AK, NH, CT, DE, GU and the countries of
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K960280IVAC NEEDLE FREE VALVE ADMINISTRATION SETS
Y, Apr 4, 1996