ClearanceBrief
Recall Z-2436-2021 ยท NEY System, Ablation, Microwave And Accessories

Angiodynamics, Inc.

Solero Generator Refurbished US PG, Item No. H78712740000USD0, Catalog No. 12740000USD

DateJul 20, 2021
StatusTerminated
ReasonSpecific serial numbers of the Solero MTA Generator require servicing to upgrade the software to help reduce the incidence of Error 0001, which can occur during system start-up.
Root causeSoftware design
ActionOn July 20, 2021, the firm initiated the recall by issuing Urgent Voluntary Medical Device Recall letters to affected consignees. Customers were informed of the software upgrade. Customers were asked to return their generators to AngioDynamics for service. The firm's Hardware Service Department will contact you to schedule the upgrade, which may include the shipment of a replacement generator to your facility if needed to be used during the upgrade of your devices. In the interim, while this upgrade is being scheduled, you may continue to use the Solero generator and are instructed to adhere
Quantity16 US, 2 OUS
DistributionDomestic distribution Nationwide. Foreign distribution worldwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K162449Solero MTA System, Solero Microwave Tissue Ablation Applicator (14cm), Solero Microwave Tissue Ablation Applicator (19cm
AngioDynamics, Inc., May 5, 2017