Recall Z-2426-2021 ยท MOI System, Test, Immunological, Antigen, Tumor

Siemens Healthcare Diagnostics, Inc

Atellica IM BR 27.29 (BR) Assay 250 Test Kit (Reference)- in vitro diagnostic use in the quantitative serial determination of cancer antigen CA 27.29 in human serum and plasma (EDTA) using the Atellica IM Analyzer.- SMN: 11206236

DateJul 6, 2021
StatusOpen, Classified
ReasonProduct does not meet the Instructions for Use (IFU) claimed Pack Calibration Interval of 10 days. IFU calibration updated interval from 10 days to 3 days for all current and future kit lot numbers
Root causeUnder Investigation by firm
ActionSiemens Healthcare issued Urgent Medical Device Correction (UMDC) via FedEx beginning on 07/06/2021. An Urgent Field Safety Notice (UFSN) was issued to all OUS Siemens sites with affected customers electronically for distribution and regional reporting to their local competent authorities beginning on 07/06/2021. Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director. Change the BR assay Pack Calibration Interval on each Atellica IM analyzer from 10 days to 3 days. See Additional Information for instructions on changing the Pac
Quantity787 units
DistributionWorldwide distribution. US Nationwide, Australia, Bahamas, India, Mexico, Saudi Arabia, and Viet Nam
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982680CHIRON DIAGNOSTICS CORPORATION ACS: CENTAUR BR IMMUNOASSAY
Chiron Diagnostics Corp., Oct 15, 1998