ClearanceBrief
Recall Z-2348-2021 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

DePuy Orthopaedics, Inc.

ATTUNE REVISION CEMENTED STEM 16X80MM-intended for use in total knee arthroplasty Part Number: 1512-16-080

DateJul 1, 2021
StatusTerminated
ReasonPackage labeled as 16mm x 80mm contained a 16mm x 130mm stem instead and result in obtaining a replacement
Root causeProcess change control
ActionDePuy Syntheses issued Medical Device Product Recall (Removal) Ref. 1986537 dated 7/1/21. Letter states reason for recall, health risk and action to take: Please take the Following Steps: 1.Examine your inventory immediately to determine if you have the subject lot and quarantine theproduct. 2.Contact your DePuy Synthes Sales Consultant to coordinate the return of any affected devices or callcustomer service following the typical returns process in order to acquire a return number prior toshipping product. 3.Review, complete, sign, and return the attached business response form (page 3 of th
Quantity7 units
DistributionInternational distribution in the countries of Canada, France.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160700ATTUNE Revision Knee System
Depuy(Ireland), Jun 10, 2016