Recall Z-2404-2021 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Artis Models with SW version VD12 listed below: Artis zee floor 10094135 Artis zee ceiling 10094137 Artis zee MP 10094139 Artis Q floor 10848280 Artis Q ceiling 10848281 Artis Q biplane 10848282 Artis Q.zen floor 10848353 Artis Q.zen ceiling 10848354 Artis Q.zen biplane 10848355 Angiography systems developed for single and biplane diagnostic imaging

DateJun 30, 2021
StatusTerminated
ReasonDue to a software error, the IAS (Image Acquisition System) may sporadically fail during startup/ restart, and result in delay of clinical treatment
Root causeSoftware design
ActionSiemens Medical Solutions USA, Inc. issued a Customer Safety Advisory Notice to affected customers via AX045/21/S dated 6/30/21. Letter states reason for recall, health risk and action to take: If the BYPASS/ BYPASS FLUORO message does not disappear after approximately four minutes, an additional restart of the Artis system is necessary. Please follow the standard emergency procedures you have in place in the event of system failures. It is recommended to consider reinforcing these emergency procedures until the update has been completed. Siemens will provide a software update for all aff
Quantity70 US (172 WW)
DistributionUS Nationwide distribution in the states of AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KY, MA, MD, MI, MN, MO, ND, NE, NH, NJ, NV, NY, OH, PA, TN, TX, UT, VA, WA, WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181407Artis zee/zeego & Artis Q/Q.zen
Siemens Medical Solution USA, Inc., Aug 15, 2018