ClearanceBrief
Recall Z-2245-2021 ยท PEN Gram-Negative Bacteria And Associated Resistance Markers

BioFire Diagnostics, LLC

FilmArray Blood Culture Identification (BCID) Panel Product Pouch Label: FilmArray BCID Panel Part No: RFIT-ASY-0126 Part No: RFIT-ASY-0127

DateJul 20, 2021
StatusTerminated
ReasonDue to product complaints received for false negative results (dropouts) with blood culture identification panel potential caused by a pouch manufacturing error.
Root causeProcess control
ActionOn 07/20/2021, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Notification Letter to customer informing them that an issue has been identified in the manufacturing process which might contribute to elevated rates of false negative results while using the BCID Panel. Actions to be taken by customers: 1) Immediately examine their inventory for product identified in this recall, 2) Discontinue use and discard any remaining product in their possession that is subject to this recall. The Recalling Firm will replace the product at no charge in accordance with its standard limited warranty. 3)
Quantity29 kits
DistributionU.S. Nationwide Distribution in the states of : FL, MI, NJ, OH, VA and WI. O.U.S.: None
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181493FilmArray Blood Culture Identification (BCID) Panel
Biofire Diagnostics, LLC, Jul 5, 2018