ClearanceBrief
Recall Z-2471-2021 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Siemens Medical Solutions USA, Inc.

ACUSON Juniper Diagnostic Ultrasound System

DateJul 12, 2021
StatusTerminated
ReasonDue to intermittent failures of the power supply in the ultrasound system which renders it inoperable.
Root causeSoftware change control
ActionOn 07/12/2021, the firm sent a "MEDICAL DEVICE SAFETY CORRECTION" Notification via email to customers informing them that there are intermittent failures of the power supply in the ultrasound system which may render the system inoperable. When an ultrasound system experiences this failure, cycling the power does not recover system functionality. Potential health risks include, lack of ultrasound guidance during interventional procedures (e.g., biopsy, guidance during surgery, patient under sedation) that rely on real-time ultrasound imaging to guide or complete the procedure. Customer will
Quantity15 systems
DistributionWorldwide distribution - U.S. Nationwide distribution in the Puerto Rico and the countries of Bosnia Herzegovina, Canada, Colombia, Germany, Greece, Italy, North Macedonia, Poland, and Slovenia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K201130ACUSON Juniper Diagnostic Ultrasound System
Siemens Medical Solutions USA, Inc., May 26, 2020