Recall Z-2384-2021 ยท LXC Monitor, Pressure, Intracompartmental

MY01, INC.

MY01 Continuous Compartmental Pressure Monitor, Reference Number MY01-0001

DateJun 28, 2021
StatusTerminated
ReasonThere is an incorrect version of the needle in the introducer part of devices. The affected Lots were manufactured using needles with a needle trocar geometry which is out of specification and may require a slightly higher push force from the user to penetrate the patient skin.
Root causeUnder Investigation by firm
ActionA Medical Device Advisory Notice dated 6/28/21 was sent to customers. Clinicians Responsibilities/ Actions required: 1. Review this notification for awareness of the contents 2. Identify and quarantine any unused affected products in your inventory for return to MY01 Inc. Contact Customer Success team by phone (+1 (855) 292-6901) or email (help@my01.io) to arrange for your replacements. 3. There are no specific patient monitoring instructions related to this voluntary product removal that are recommended beyond your existing follow up schedule. 4. Complete Attachment 1- Certificate of acknowl
Quantity37 units
DistributionUS Nationwide distribution in the states of MD, OH, CA, NY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K210525MY01 Continuous Compartmental Pressure Monitor
MY01, Inc., Apr 22, 2021