Recall Z-2357-2021 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Philips Medical Systems Gmbh, DMC

CombiDiagnost R90 Software Version R1.0 and R1.1

DateJul 22, 2020
StatusOpen, Classified
ReasonPhilips has become aware of a software issue, where the kV and mA values are not locked immediately but change when the collimator shutter position is changed directly after the Lock-In command. This software issue leads to an over or under exposed image.
Root causeRadiation Control for Health and Safety Act
ActionPhilips sent customers an Electronic Product Radiation Correction letter dated July 14, 2021. The letter identified the affected product, problem and actions to be taken. Philips plans to correct the defect at no cost to customers by implementing a field correction involving the installation of a software update. For questions contact the Customer Care Solutions Center 1-800-722-9377 and reference FCO70900051.
Quantity322 systems; 66 in US
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K203087CombiDiagnost R90
Philips Medical Systems Dmc GmbH, Dec 3, 2020