Recall Z-2333-2021 ยท MVU Reagents, Specific, Analyte

Becton, Dickinson and Company, BD Biosciences

BD" Mouse IgG2a Isotype Control APC X39 ASR

DateJul 22, 2021
StatusTerminated
ReasonComplaints have been received for performance issues related to high background and staining. this issues is related to a manufacturing process error that can potentially cause a false positive staining due to contamination with another antibody.
Root causeNonconforming Material/Component
ActionOn 07/22/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification letter via FedEx to customers to inform them that the Recalling Firm received 3 complaints for performance issues related to high background and staining. Our investigation has confirmed there was an error during the manufacturing process that can potentially cause a false positive staining due to contamination with another antibody. Customers are instructed to: 1) Immediately review their inventory for the specific Catalog and lot number listed in the Recall Notification above. Destroy all product subject to the reca
Quantity23 reagents
DistributionWorldwide distribution - U.S. Nationwide distribution in the states of CT, FL, KY, NY, and PA. The country of Taiwan.