ClearanceBrief
Recall Z-2315-2021 ยท JAK System, X-Ray, Tomography, Computed

Philips North America Llc

Model 728321/728323 CT5000 Ingenuity Plus/Pro/Premium-Computed Tomography X-ray system

DateJun 8, 2021
StatusTerminated
ReasonWhen setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.
Root causeSoftware design
ActionPhilips issued Urgent Medical Device Correction letter dated 6/8/21 to consignees via certified mail. Letter states reason for recall, health risk and action to take: following steps: Go to Preferences> Regional Settings > Weight units, select the Weight units to Kg. Step 1: Verify that the patient weight unit is set to Kg, rather than Lbs. Step 2: Change the weight unit to Kg if weight unit is set to Lbs, referring to the screenshot below Step 3: Log out and then log back in to make the changes effective. Please share this notice with all those who need to be aware within your organization.
Quantity2 units
DistributionUS Nationwide distribution in the states of AZ, FL, LA, MN, NE, NV, NY, PA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160743Philips Ingenuity CT
Philips Medical Systems (Cleveland), Inc., Aug 8, 2016