Recall Z-2340-2021 ยท MJN Catheter, Intravascular Occluding, Temporary
W. L. Gore & Associates Inc.
Gore Molding & Occlusion Balloon Catheter, REF Catalogue Number MOB37, SN XXXXXXXX, 10 mm - 37 m 10 Fr, Guidewire Compatibility 0.035", STERILE EO, CE2797, UDI: (00)733132639489
| Date | Jul 12, 2021 |
| Status | Open, Classified |
| Reason | Complaints received concerning Balloon Catheter leakage from the guidewire lumen and the y-hub during preparation resulting in either insufficient inflation or an inability to inflate. |
| Root cause | Process change control |
| Action | On July 23, 2021 issue a "Urgent Medical Device Recall" notification to affected consignees via UPS. In addition to informing consignees about the recall the customer notification ask consignees to take the following actions: 1. Identify and return any unused devices within the scope of this voluntary recall. 2. Please complete and sign the enclosed CUSTOMER RESPONSE FORM and return to WLGore4731@stericycle.com within 2 weeks of receipt of this notification. 3. Please share this letter with others in your hospital or clinic as appropriate. 4. If a listed device has been used, there is no |
| Quantity | 3,364 units |
| Distribution | Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of AUS, CAN, CHE, DE |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K172567 | GORE Molding and Occlusion Balloon Catheter W. L. Gore and Associates, Inc., Jan 31, 2018 |