Recall Z-2306-2021 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic
Philips Ultrasound Inc
EPIQ Ultrasound Systems, Model: EPIQ Elite, EPIQ 5G, EPIQ 5C, EPIQ 5W, EPIQ 7G, EPIQ 7C, EPIQ 7W, EPIQ CVx, EPIQ CVxi
| Date | Jul 22, 2021 |
| Status | Completed |
| Reason | Ultrasound system control panel arm swivel lock could fail, which could cause the control panel and monitor to swivel freely while a user is positioning the monitor, or transporting/moving the system, which could lead to patient or user musculoskeletal injury. |
| Root cause | Device Design |
| Action | Medical device correction notices were mailed to customers who were asked to take the following actions to prevent risks for patients or users: When Positioning or Transporting/Moving the EPIQ Ultrasound System: A) Prior to positioning or transporting/moving the system, ensure the control panel locking mechanism is engaged to prevent the control panel and monitor from swiveling freely. B) To ensure the control panel locking mechanism is working properly: 1. Engage the system brake. 2. Observe the lock/unlock button on the system handle. It should not be illuminated. 3. If it is illuminated pre |
| Quantity | 8935 |
| Distribution | Worldwide distribution - US Nationwide distribution including in the states of FL, UT, CA, CO, WA, MS, NC, NV, GA, LA, VA, SC, NE, TN, HI, OR, ID, MI, OH, IN, IL, OK, TX, PA, NJ, CT, NY, DE, SD, WI, VT, MN, KS, MO, KY, MA, NH, IA, MT, NM, ME, AZ, AR, AL, WY, ND, AK, MD, RI, WV, DC, VI and the countr |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K182857 | EPIQ Diagnostic Ultrasound System, EPIQ 5 Diagnostic Ultrasound System, EPIQ 7 Diagnostic Ultrasound System, Affiniti 30 Philips Ultrasound, Inc., Nov 9, 2018 |