ClearanceBrief
Recall Z-2259-2021 ยท KWA Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)

Biomet, Inc.

Arcos Modular Revision Hip System Standard Cone Prox Body, Size A, 50 MM Item Number: 11-301300

DateJun 2, 2021
StatusTerminated
ReasonArcos Proximal Cone Bodies could potentially exhibit chatter in the inner taper fretting corrosion and device failure in vivo leading to surgical intervention
Root causeProcess control
ActionZimmer Biomet issued Urgent Medical Device Notification Letter on 6/2/21 to Distributors/ Risk Managers/Surgeons, stated reason for recall, health risk and action to take: Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the
Quantity88 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090757BIOMET MODULAR FEMORAL REVISION SYSTEM
Biomet, Inc., Sep 25, 2009