Recall Z-2158-2021 ยท INM Chair, With Casters

Sunrise Medical (US) LLC

Zippie Voyage, early intervention stroller - Product Usage: is manually operated by a caregiver for purposes of assisting a disabled person while providing supportive elements for assisted transportation and seating for a child or small adult not to exceed 75 pounds.

DateJul 8, 2021
StatusTerminated
ReasonThe seating system unexpectedly detached, which resulted in the seating system falling to the ground, causing injury.
Root causeDevice Design
ActionOn 07/12/2021, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Notification informing customers that if the seating system is not adequately attached to the stroller base, the seating system may slide off the base and fall to the ground potentially causing injury a child. The Recalling Firm will be providing a safety tether and are informing/instructing their customers to: This Safety Tether Kit will be provided free of charge to all consumers starting the week of July 12, 2021. The Safety Tether Kit will be accompanied by instructions to guide consumers through installing and using the safe
Quantity4,741 device
DistributionWorldwide distribution - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. Th
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.