ClearanceBrief
Recall Z-2257-2021 ยท QKP Coronavirus Antigen Detection Test System.

Beckman Coulter Inc.

ACCESS SARS-CoV-2 ANTIGEN QC, REF: C68670, CONTENTS: 3 X 6.0 mL QC1, 3 x 6.0 mL QC2, 1 QC VALUE CARD, IVD, Rx Only, UDI: (01)15099590742737;

DateJul 1, 2021
StatusTerminated
ReasonSARS-CoV-2 Antigen Assay test may generate false positive results which could result in incorrect diagnosis .
Root causeOther
ActionOn 07/09/2021 Beckman Coulter issued an "URGENT MEDICAL DEVICE RECALL" to all affected consignees via email and US Postal Mail. In addition to informing consignees about the recalled device, the firm ask consignees to take the following actions: 1. Immediately discontinue using the Access SARS-CoV-2 Antigen assay. 2. Discard all Access SARS-CoV-2 Antigen reagent packs. 3. At the discretion of the Laboratory Medical Director, consider updating previous laboratory reports to indicate prior Reactive laboratory results may not be accurate. 4. Please share this information with your laboratory s
Quantity544 units
DistributionUS Nationwide Distribution in the states of Alabama Alaska Arizona Arkansas California Connecticut Delaware Florida Georgia Idaho Illinois Indiana Iowa Kentucky Louisiana Maine Maryland Massachusetts Michigan Mississippi New Hampshire New Jersey New York North Carolina Ohio Oklahoma Oregon Pennsylv
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.