Recall Z-2140-2021 ยท DWC Controller, Temperature, Cardiopulmonary Bypass

Maquet Medical Systems USA

Heater Cooler Unit (HCU 30) The device supplies temperature-controlled water for cardioplegia heat exchangers and for blood heat exchangers in extracorporeal circulation and for blankets with which patients can be warmed or cooled.

DateJul 14, 2021
StatusTerminated
ReasonThere is the potential for contaminated water to enter parts of the Heater-Cooler Unit circuits. There is a potential risk of bacterial contamination including Mycobacteria species.
Root causeDevice Design
ActionOn July 14, 2021, Maquet Getinge Group issued Urgent Medical Device Recall notices to customers via FedEx 2 day Delivery with Signature Proof of Delivery. Customers were advised to take the following action: - Users should take the unit out of operation at the earliest opportunity and decommission the unit. - MCP has not developed a cleaning protocol that meets current concerns and expectations of the FDA. - If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. - Please acknowledge receipt of t
Quantity33
DistributionDistribution to US states of AL, CA, FL, IL, KS, MI, MO, NC, NE & OH.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031544HEATER COOLER UNIT, MODEL HCU 30
Jostra AB, Aug 7, 2003