Recall Z-2278-2021 ยท DQE Catheter, Oximeter, Fiber-Optic

Edwards Lifesciences, LLC

Edwards PediaSat Oximetry Catheter Kit, REF: XT3515KTP, Exterior Diameter 5.5F (1.83 mm), Usable Length 15 cm, Number of Lumens 3, Recommended Guidewire Size 0.025" (0.64 mm), Rx Only, Sterile EO, UDI: (01)00690103197563

DateJul 9, 2021
StatusTerminated
ReasonThere is a potential for internal leaks within catheters.
Root causeProcess control
ActionOn 7/9/2021 Edwards issued a "URGENT MEDICAL DEVICE PROUDCT CORRECTION" via FedEx to all affected consignees. In addition to informing consignees about the recall the firm ask consignees take the following actions: 1. The problem may be identified during preparation of the catheter for insertion per the IFU, paying specific attention to the connection between the green hub and the tubing for any signs of leakage. Additionally, the problem may be identified if when flushing the lumens of the catheter while administering a medication, a transient change of vital signs is observed. In either o
Quantity70 UNITS
DistributionUS: CA, TX, FL, AZ, PA, IN, NY, NJ, MI, WA, MD, OH, FL, OUS: Japan, Mexico, Brazil, Colombia, Taiwan, India, Poland, Germany, United Arab Emirates, Israel, Switzerland, Bahrain, Iman, Tunisia, Austria, Belgium, France, Netherlands, Italy, Spain, Canary Islands, Portugal, Greece, Costa Rica, Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K053609PRESEP AND PEDIASAT OXIMETRY CATHETERS AND VIGILEO ARTERIAL PRESSURE CARDIAC OUTPUT/OXIMETRY MONITOR
Edwards Lifesciences, LLC, Feb 28, 2006