Recall Z-2244-2021 ยท FOS Catheter, Umbilical Artery

Cardinal Health 200, LLC

Argyle UVC Insertion Tray containing Safety Scalpel N11 (vendor part number 73-1811). Intended for use to insert an Umbilical Venous Catheter (UVC) into the umbilical artery or vein of neonate. Product Code: 43201

DateAug 13, 2021 (initiated Jun 16, 2021)
StatusTerminated
ReasonThe current Argyle UVC Insertion Tray does not include a specific IFU to operate the Safety Scalpel N11. The Safety Scalpel N11 features a permanent locking feature that will not allow the clinician to unlock the scalpel once the safety shield is closed and pushed forward into a locked position may result in a delay treatment/therapy, which could ultimately result in death.
Root causeLabeling Change Control
ActionCardinal Health issued Urgent Medical Device Correction letter and Acknowledgement Form on June 16, 2021 to the direct consignees. Letter states reason for recall, health risk and action to take: Actions Required: 1) REVIEW the IFU for the Safety Scalpel N11 (vendor part number 73-1811). 2) COMMUNICATE the IFU to all personnel that utilize the Argyle" UVC Insertion Tray (SKU 43201) containing Safety Scalpel N11 (vendor part number 73-1811) 3) NOTIFY any customers to whom you may have distributed/forwarded affected product or will send the product about this product correction notice and shar
Quantity7197 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.