Recall Z-2365-2021 ยท MSD Catheter, Hemodialysis, Implanted
Covidien, LP
Palindrome RT Repair Kit 15 FR TAL 23cm, Sterile, Item Code 8888541123
| Date | May 28, 2021 |
| Status | Open, Classified |
| Reason | This voluntary withdrawal is being conducted to implement point-of-use labeling on each device and enhancements to the Instructions for Use (IFU). |
| Root cause | Labeling design |
| Action | A Medical Device Market Withdrawal notification letter dated 5/28/21 was sent to customers. Actions being taken by Medtronic 1. Medtronic is developing point-of-use labeling reinforcing the important user assembly steps contained within the IFU related to the hub/back end assembly procedure. 2. The IFU for Palindrome" RT and Palindrome" Precision RT chronic hemodialysis catheters will be updated to enhance the detailed instructions for users related to the steps required for hub/back end assembly. Actions you should take 1. Please immediately quarantine and discontinue use of affected item c |
| Distribution | Worldwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K111817 | PALINDROME RT CHRONIC CATHETER Covidien, Aug 30, 2011 |