ClearanceBrief
Recall Z-2363-2021 ยท MSD Catheter, Hemodialysis, Implanted

Covidien, LP

Palindrome RT Chronic Catheter Kit Symmetrical Tip, Reverse-Tunneled 15 Fr/Ch (5.0 mm) x 55cm, Item Code 8888541055

DateMay 28, 2021
StatusOpen, Classified
ReasonThis voluntary withdrawal is being conducted to implement point-of-use labeling on each device and enhancements to the Instructions for Use (IFU).
Root causeLabeling design
ActionA Medical Device Market Withdrawal notification letter dated 5/28/21 was sent to customers. Actions being taken by Medtronic 1. Medtronic is developing point-of-use labeling reinforcing the important user assembly steps contained within the IFU related to the hub/back end assembly procedure. 2. The IFU for Palindrome" RT and Palindrome" Precision RT chronic hemodialysis catheters will be updated to enhance the detailed instructions for users related to the steps required for hub/back end assembly. Actions you should take 1. Please immediately quarantine and discontinue use of affected item c
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111817PALINDROME RT CHRONIC CATHETER
Covidien, Aug 30, 2011