Recall Z-2138-2021 ยท MUJ System, Planning, Radiation Therapy Treatment

RAYSEARCH LABORATORIES AB

RayStation (Radiation Treatment Planning System) : RayStation 9A, 9B, 9B SP1, 10A, 10A SP1, 10B and 11A, in combination with RayCare Model Numbers/ UDI: RayStation 9.0 (9.0.0.113) 07350002010174; RayStation 9.1 (9.1.0.933) 07350002010266; RayStation 9.1 Service Pack 1 (9.2.0.483) 07350002010297; RayStation 10.0 (10.0.0.1154) 07350002010303; RayStation 10.0 Service Pack 1 (10.0.1.52) 0735000201

DateJun 8, 2021
StatusTerminated
ReasonInitial delivery positions will be set incorrectly when having setup beam(s) at the localization point.
Root causeUnder Investigation by firm
ActionRaySearch issued Urgent Field Safety Notice, Medical Device Correction #80230 via email on 06/08/21. Letter states reason for recall, health risk and action to take: TO BE TAKEN BY THE USER " Do not create setup beams with the localization point as isocenter (option 2). " Always perform secondary verification of new treatment positions. RayTreat will notify the user when the treatment positions have been updated. " Always perform imaging before treating the patient. " Educate planning staff and all users about this workaround. " Inspect your product and identify all installed units with the a
Quantity5 units
DistributionUS distribution to TN; and Belgium
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K200487RayCare 3.1
RaySearch Laboratories AB (PUBL), Jun 10, 2020
K200569RayStation
RaySearch Laboratories AB (PUBL), Sep 24, 2020