Recall Z-2460-2021 ยท FPA Set, Administration, Intravascular

CAREFUSION

BD Extension Set SMALLBORE Tubing SmartSite VALVE, REF: 10010511; BD SmartSite Extension Set Needle-Free Valve, REF: 10010912; BD EXTENSION SET SMALLBORE TUBING DETACHABLE SmartSite VALVE, REF: 10010983; BD Gravity Burette Set 60 DP 4 VersaSafe PORTS 3 SmartSite VALVES BALL VALVE DRIP CHAMBER V/NV, REF: 10011087; BD EXTENSION SET SMALLBORE TUBING SmartSite VALVE (DETACHABLE), REF: 10011253; BD EX

DateAug 2, 2021
StatusOpen, Classified
ReasonNeedle-free connector valves, either standalone or attached to extension sets, may have issues including difficulty to flush, flow issues, and partial or total occlusions, which could lead to therapy delay. Follow-up activates not required if product has already been used or was safely connected to the female luer of connecting device.
Root causeProcess control
ActionOn 08/02/21, Urgent Medical Device Correction notices were mailed and emailed to consignees who were asked to notify recalling firm if this issue was experienced at 1-844-823-5433, productcomplaints@bd.com . Follow-up activates not required if product has already been used or was safely connected to the female luer of connecting device. If during priming the clinician encounters an occluded needle-free connector, take the following actions: 1) Depress the syringe plunger while the syringe is still attached to the needle-free connector. 2) If unable to depress the plunger, pull back on the syri
Quantity20,874,700
DistributionWorldwide distribution - US Nationwide distribution in the states of FL, PA, NC, OH, SC, IL, RI, MO, NY, TN, AL, VA, PR, CA, NE, NJ, MI, MD, NM, LA, KY, NV, ME, OK, WA, MA, WV, TX, AR, OR, KS, GA, VT, IN, WY, AZ, CO, MN, WI, UT, SD, MS, ID, MT, ND, HI, IA, DC, AK, NH, CT, DE, GU and the countries of
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K960280IVAC NEEDLE FREE VALVE ADMINISTRATION SETS
Y, Apr 4, 1996