Recall Z-2294-2021 ยท GEX Powered Laser Surgical Instrument

Olympus Corporation of the Americas

Soltive Premium Super Pulsed Laser System (TFL-PLS) & Pro Super Pulsed Laser System (TFL-SLS). Intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in urology, lithotripsy, gastroenterological surgery and gynecological surgery

DateJun 30, 2021
StatusTerminated
ReasonThermal injury following dusting and fragmenting treatment of ureteral stones when user exceeded the 20W standard presets
Root causeSoftware design
ActionOlympus issued Urgent Medical Device Correction on June 30, 2021 via Federal Express Second Day Delivery. Letter states reason for recall, health risk and action to take: To mitigate and prevent thermal injury events, Olympus is issuing a software upgrade for the Soltive Laser to version 2.1. Software version 2.1 includes a new Ureteral Stone Preset of 8W under the Lithotripsy treatment mode. The Soltive Laser Instructions for Use (IFU) has been updated to reflect these software changes which also includes a pop-up alert reminding user to consider the potential clinical consequences of se
Quantity408 units
DistributionUS Nationwide Distribution Foreign: Canada Australia Japan Hong Kong Europe Singapore
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K211401SOLTIVE Laser System (SOLTIVE Pro SuperPulsed Laser, SOLTIVE Premium SuperPulsed Laser, SOLTIVE Laser Fibers, and Access
Gyrus Acmi, Inc., Jun 3, 2021