ClearanceBrief
Recall Z-2173-2021 ยท FKX System, Peritoneal, Automatic Delivery

BAXTER HEALTHCARE CORPORATION

Sharesource Connectivity Platform for Use with the Amia Automated PD System

DateJun 30, 2021
StatusTerminated
ReasonBaxter Healthcare Corporation has identified that the AMIA Sharesource User Guide that is used with the AMIA Automated PD System cyclers incorrectly describes the function of the UF (Ultrafiltration) Limit setting as the ending criteria of the cyclers Extra Last Drain Mode. In some locations in the guide, it states that the UF volume achieved must exceed the programmed UF Limit before the cycler will end the Extra Last Drain, allowing the patient to end therapy or proceed to their last fill.
Root causeUnder Investigation by firm
ActionThe firm notified its consignees by letter on 06/30/2021. The letter explained the issue with the instructions for use, specifically, That the AMIA Sharesource User Guide describes the function of the UF (ultrafiltration) Limit setting as the ending criteria of the cycler's Extra Last Drain Mode. Baxter will be sending a Follow-Up communication once the AMIA Sharesource User Guide has been updated on the portal.
Quantity1 unit
DistributionWorldwide distribution - US Nationwide distribution and the country of Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151525Amia Automated PD System with Sharesource
Deka Research and Development, Oct 9, 2015