ClearanceBrief
Recall Z-2100-2021 ยท NDN Cement, Bone, Vertebroplasty

Stryker Spine

Cortoss Bone Augmentation Material 5cc Cartridge, Catalog Number 2101-0005, GTIN 00808232000962

DateMay 18, 2021
StatusTerminated
ReasonThe products experienced a 12-hour excursion from the required refrigerated temperature conditions at the distribution center prior to distribution.
Root causeStorage
ActionUrgent Medical Device Recall notification letters dated 5/14/21 were sent to customers. Actions needed NOTE: due to the potential hazards/risks associated with use of product associated with these shipments, please urgently quarantine the in-scope product. 1. Immediately review the Product Recall Notification. Please forward this information to all appropriate personnel within your organization. Maintain awareness of this communication internally until all required actions have been completed within your facility. 2. Immediately check your internal inventory to locate the affected products.
Quantity559 total
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080108CORTOSS BONE AUGMENTATION MATERIAL
Orthovita, Inc., Jun 5, 2009