ClearanceBrief
Recall Z-2029-2021 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

ARTIS pheno- angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Model: 10849000

DateMay 27, 2021
StatusOpen, Classified
ReasonSoftware error VE10B -In case of a fault within the drivetrain of an axis and while performing a stepped peripheral imaging (PERISTEPPING/PERIVISION), the C-arm may leave its intended travel path by up to 22 cm (9") result in canceling clinical treatment or to continue treatment on an alternative system
Root causeSoftware design
ActionSiemens issued a Urgent Medical Device Correction Letter on 5/27/21 via AX064/20/S. Additionally, a software solution is being distributed via Update Instruction AX063/20/S to all affected customers. Letter states reason for recall, health risk and action to take: Planned procedures may have to be terminated and performed on an alternative X-ray system. Please follow the standard emergency procedures you have in place in the event of system failures. It is recommended to consider reinforcing these emergency procedures until the update has been completed. Siemens will update the affected so
Quantity116 units
DistributionUS Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163286ARTIS pheno
Siemens Medical Solution USA, Inc., Mar 9, 2017
K201156ARTIS pheno (VE2)
Siemens Medical Solution USA, Inc., Jun 29, 2020