Recall Z-2161-2021 ยท OWB Interventional Fluoroscopic X-Ray System

GE OEC Medical Systems, Inc

OEC Elite mobile C-arm - Product Usage: intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during diagnostic, interventional, and surgical procedures.

DateJun 10, 2021
StatusTerminated
ReasonThere is a potential that the coin cell battery used to monitor X-Ray tube temperature may deplete early causing the system to become inoperable.
Root causeUnder Investigation by firm
ActionOn June 11, 2021, GE Healthcare issued an "Urgent Medical Device Correction via certified mail to all affected consignees. In addition to informing consignees about the recalled product, the firm asked customers to take the following actions: 1. You may continue to use your system. 2. Please disseminate this information to applicable users within your facility. 3. When the system is not in use, it is recommended to connect the system to power with the workstation plugged in and the interconnect cable attached to the mainframe, to reduce potential battery depletion. It is not necessary to t
Quantity3570 units
DistributionWorldwide distribution - US Nationwide distribution including in the states of CA, PA, WA, OH, LA, MS, AL, FL, MO, OK, AZ, NJ, CO, MI, NV, MN, KS, TN, NC, IL, MD, SC,GA, UT, WI, MT, AR, WV, ID,CT, IA, WY, VT, KY, DC, SD,ME, Virgin Islands and the countries of Colombia, Egypt, Peru, Chile, Spain, Fra
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170752OEC Elite
Oec Medical Systems, Inc., May 19, 2017
K171565OEC Elite
Ge Oec Medical Systems, Inc., Aug 10, 2017
K172550OEC Elite
Ge Oec Medical Systems, Inc., Nov 16, 2017
K192819OEC Elite
Ge Oec Medical Systems, Inc., Nov 8, 2019