Recall Z-2160-2021 ยท OXO Image-Intensified Fluoroscopic X-Ray System, Mobile

GE OEC Medical Systems, Inc

OEC 3D Imaging, Image-Intensified Fluoroscopic X-Ray System, Mobile - Product Usage: designed to provide fluoroscopic and digital spot images of adult and pediatric populations during diagnostic, interventional, and surgical procedures

DateJun 10, 2021
StatusTerminated
ReasonThere is a potential that the coin cell battery used to monitor X-Ray tube temperature may deplete early causing the system to become inoperable.
Root causeUnder Investigation by firm
ActionOn June 11, 2021, GE Healthcare issued an "Urgent Medical Device Correction via certified mail to all affected consignees. In addition to informing consignees about the recalled product, the firm asked customers to take the following actions: 1. You may continue to use your system. 2. Please disseminate this information to applicable users within your facility. 3. When the system is not in use, it is recommended to connect the system to power with the workstation plugged in and the interconnect cable attached to the mainframe, to reduce potential battery depletion. It is not necessary to t
Quantity3 units
DistributionWorldwide distribution - US Nationwide distribution including in the states of CA, PA, WA, OH, LA, MS, AL, FL, MO, OK, AZ, NJ, CO, MI, NV, MN, KS, TN, NC, IL, MD, SC,GA, UT, WI, MT, AR, WV, ID,CT, IA, WY, VT, KY, DC, SD,ME, Virgin Islands and the countries of Colombia, Egypt, Peru, Chile, Spain, Fra
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K203346OEC 3D
Ge Oec Medical Systems, Inc., Mar 5, 2021