Recall Z-2166-2021 ยท PTM Syringe, Balloon Inflation, Exempt

Merit Medical Systems, Inc.

basixALPHA Inflation Device 30 atm/bar 20 mL - Product Usage: Used to inflate and deflate an angioplasty balloon or other interventional device, and to measure the pressure within the balloon.

DateJun 10, 2021
StatusTerminated
ReasonDue to a design issue, inflation syringe handle separating from the inflation device when handle is withdrawn.
Root causeDevice Design
ActionOn 06/08/2021, the firm sent an "URGENT PRODUCT RECALL NOTICE" initially via an email communication and then followed up with a hardcopy letter via FedEx overnight on 06/10/2021. The communication and letter informed customers that due to a design defect the handle of the inflation device may separate from the syringe during use, which may result in a delay in procedure. Customers are instructed to: 1. Please immediately determine if any of the devices identified in the attached Customer Response Form (CRF) are within their facility, quarantine them, and discontinue use and distribution. 2.
Quantity7,211 devices
DistributionWorldwide distribution - U.S. Nationwide distribution in the states of AL, AZ, CA, CO, DE, FL, GA, ID, IL IN, KS, LA, MI, MN, MO, NC, NE, NJ, NV, NY, OH, OR, PA, SC, TN, TX, UT, WA, WI, WV, and WY. The countries of Australia, Austria, China, Denmark, Finland, France, Germany, Israel, Japan, Kenya, K
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.