Recall Z-2135-2021 ยท MIH System, Endovascular Graft, Aortic Aneurysm Treatment

Medtronic Vascular, Inc.

Endurant II/Endurant IIs Stent Graft System: Medtronic Endurant II Stent Graft System REF ETUF3214C102EE, ETCF3636C49EE, ETBF3216C166EE, and ETUF3614C102E. Used to treat infrarenal abdominal aortic or aortoiliac aneurysms using an endovascular approach.

DateJun 7, 2021
StatusOpen, Classified
ReasonDuring stent graft deployment, the radiopaque (RO) marker bond detached from the distal end of the graft cover.
Root causeUnder Investigation by firm
ActionOn 06/07/2021, the firm initiated a verbal notification regarding an "URGENT MEDICAL DEVICE RECALL" which informed customers that Earlier this week, Medtronic identified the potential for a specific subset of 20Fr Endurant II/IIs Stent Graft Systems to have the potential for the Radiopaque Marker to detach during deployment. There is evidence that devices that used one batch of graft cover subassemblies has a higher potential to exhibit this failure mode. This is an acute failure of the delivery system, there are no concerns with previously implanted stent grafts. We are requesting that you im
Quantity18 devices
DistributionU.S.: GA, MA, MD, NY, SC, TX, WA, and WV O.U.S.: Argentina, Australia, Germany, Greece, Israel, Italy, Netherlands, Poland, South Africa, Spain, Switzerland and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.