ClearanceBrief
Recall Z-2010-2021 ยท OCC Respiratory Virus Panel Nucleic Acid Assay System

Luminex Corporation

Verigene Clostridium difficile Nucleic Acid Test Kit (CDF) Model Number: 20-006-022

DateOct 30, 2020
StatusOpen, Classified
ReasonPotential for a false-negative result.
Root causeDevice Design
ActionOn October 2020 the firm sent a letter to all their consignees with the following information: Dear Valued Luminex Customer, According to our records, you have ordered and may have used our portfolio of VERIGENE Assays in your laboratory. Through post-market surveillance activities, we have identified an increase in the No Call/False Positivity rate for these products. Our investigation has identified that the likelihood of a No Call/False Positive for the assays below increases when test cartridges are left on the Processor SP for an extended time following completion of assay preparation.
Quantity1,261
DistributionWorldwide distribution - US Nationwide including in the states of AL, AZ, CA, CO, CT, DE, DC, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, NE, NV, NH, NJ, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA and VT. The countries of Austria, Chile, France, Germany, Greece, Italy, Ja
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123197VERIGENE C. DIFFICILE NUCLEIC ACID TEST (CDF)
Nanosphere, Inc., Dec 5, 2012