Recall Z-2007-2021 ยท OCC Respiratory Virus Panel Nucleic Acid Assay System
Luminex Corporation
Verigene Gram-Positive Blood Culture Nucleic Acid Test (BC-GP) Model Number: 20-006-018
| Date | Oct 30, 2020 |
| Status | Open, Classified |
| Reason | Potential for a false-negative result. |
| Root cause | Device Design |
| Action | On October 2020 the firm sent a letter to all their consignees with the following information: Dear Valued Luminex Customer, According to our records, you have ordered and may have used our portfolio of VERIGENE Assays in your laboratory. Through post-market surveillance activities, we have identified an increase in the No Call/False Positivity rate for these products. Our investigation has identified that the likelihood of a No Call/False Positive for the assays below increases when test cartridges are left on the Processor SP for an extended time following completion of assay preparation. |
| Quantity | 36,416 |
| Distribution | Worldwide distribution - US Nationwide including in the states of AL, AZ, CA, CO, CT, DE, DC, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, NE, NV, NH, NJ, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA and VT. The countries of Austria, Chile, France, Germany, Greece, Italy, Ja |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K122514 | VERIGENE GRAM-POSITIVE BLOOD CULTURE (GC-GP) NUCLEIC ACID TEST Nanosphere, Inc., Sep 19, 2012 |