Recall Z-2053-2021 ยท LNH System, Nuclear Magnetic Resonance Imaging
GE Healthcare, LLC
SIGNA Architect - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
| Date | Jul 12, 2021 (initiated Jun 4, 2021) |
| Status | Open, Classified |
| Reason | GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation on the system prior to running the initial 3-Plane Localizer Scan (See Figure 1 for reference to buttons.) Selecting and saving a patient orientation that does not match the patient's actual position may result in incorrectly annotated and/or flipped images. |
| Root cause | Software design |
| Action | Deployment of the original customer letter to consignees in the U.S. began on June 4, 2021. Dissemination of the updated customer letter to consignees in the U.S. began on June 11, 2021. The letters discuss an issue on the affected products listed below where the Patient Orientation button may inadvertently be clicked when intending to click on the Save RX button. Users are directed to review and confirm that the actual patient orientation matches the prescribed patient orientation before initiating the scan. A correction is planned. |
| Quantity | 311 devices |
| Distribution | Worldwide distributions. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K160618 | Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T, Optima MR450w 1.5T Ge Medical Systems, LLC, Jun 9, 2016 |
| K163331 | Discovery MR750 3.0T; Discovery MR750w 3.0T;Discovery MR450 1.5T; Discovery MR450w 1.5T; SIGNA Architect and SIGNA Artis Ge Medical Systems, LLC, Mar 17, 2017 |
| K202966 | SIGNA Architect Ge Healthcare (Ge Medical Systems, LLC), Nov 13, 2020 |