Recall Z-2072-2021 ยท DSQ Ventricular (Assist) Bypass

Heartware, Inc.

HeartWare, HVAD Pump Implant Kit, REF MCS1705PU - Product Usage: indicated for hemodynamic support in patients with advanced, refractory left ventricular heart failure; either as a Bridge to Cardiac Transplantation (BTT), myocardial recovery, or as Destination Therapy (DT) in patients for whom subsequent transplantation is not planned.

DateMay 28, 2021
StatusTerminated
ReasonThe pump has an impeller with a shroud height that did not meet the lower control limit.
Root causeNonconforming Material/Component
ActionThe firm initiated the recall to the sole consignee on 05/28/2021 by telephone. The consignee was asked to quarantine and return the affected device.
Quantity1 unit
DistributionInternational distribution to the country of Germany.