Recall Z-2042-2021 ยท NPT Aortic Valve, Prosthesis, Percutaneously Delivered

Medtronic Heart Valves Division

CoreValve Evolut PRO Delivery Catheter System, REF: ENVPRO-14; ENVPRO-16;

DateJun 11, 2021
StatusOpen, Classified
ReasonDue to reports of actuator separation for various delivery catheter systems. Affected products were manufactured before July 11, 2020.
Root causeDevice Design
ActionOn 06/09/2021, the Recalling Firm sent an "URGENT MEDICAL DEVICE RECALL" Notification via UPS 2-day (US Customers) and regionally approved methods (OUS Customer) to customers informing them that it has been identified that were is the potential for actuator separation with delivery catheter systems manufactured before 07/11/2020. Customer are instructed to: 1. Identify and quarantine all unused affected product manufactured before 11-JUL-2020.The enclosed Customer Notification Detail Report includes affected product for their facility. 2. Return all unused affected product in their inventor
QuantityTotal of all Delivery Systems (Globally) = 67,339 units
DistributionU.S.: AL, AK, AR, AZ, CA, CT, DE, DC, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NE, NH, NJ, NM, NV, NY, NC, ND, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. O.U.S.: Albania, Argentina, Australia, Austria, Belarus, Belgium, Brazil, BULGARIA, Cana