ClearanceBrief
Recall Z-2040-2021 ยท QDS Mers-Cov And Common Respiratory Pathogens Semi-Quantitative And Quantitative Multiplex Nucleic Acid Detection System

BioFire Diagnostics, LLC

FilmArray Pneumonia Panel plus (Pneumoplus) Part Numbers: Part No: RFIT-ASY-0143 / UDI: 00815381020314 (30 test kit) Part No: RFIT-ASY-0142 / UDI: 00815381020321 (6 test kit)

DateJun 7, 2021
StatusTerminated
ReasonDue to real-time stability study failures resulting in increase rate of false negative results between the 6 to 12 month expiry time-frame.
Root causeUnder Investigation by firm
ActionOn 06/03/2021, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Notification Letter via overnight FedEx informing its customer that it has identified a potential for obtaining elevated rates of negative adenovirus results when using the pneumonia panel. The potential false negative results are due to a 10-100x reduction in sensitivity specifically for adenovirus species C when using the pneumonia panel kits within 6 months of expiration. In addition: -The test performance is NOT impacted if kits are more than 6 months from expiration date. -Performance for other adenovirus species (e.g. A
Quantity3,559 kits (Globally)
DistributionU.S.: AK, AL, AZ, CA, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, IN, KS, KY, KY, LA, MA, MD, ME, MI, MI, MN, MO, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TN, TX, UT, VA, VT, WA, WI, WV and WY. O.U.S.: Algeria, Angola, Anguilla, Argentina, Australia, Austri
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181324FilmArray Pneumonia Panel plus
Biofire Diagnostics, LLC, Nov 15, 2018