ClearanceBrief
Recall Z-2139-2021 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

GE Healthcare, LLC

CARESCAPE PDM-Masimo SpO2. Physiological data monitor.

DateJul 23, 2021 (initiated May 21, 2021)
StatusOpen, Classified
ReasonMasimo SpO2 Saturation Values can become frozen after an extended length of use without a power down.
Root causeSoftware Manufacturing/Software Deployment
ActionOn May 21st the firm sent a letter to its customers with the following instructions: You can continue to use the CARESCAPE PDM with Masimo SET technology by completing the power down instructions below annually. 1. If the PDM is on a patient, provide alternate monitoring if needed, during the power down process. 2. Disconnect the PDM from the display device by removing the PDM from the mounting dock or by disconnecting the communication cable from the back of the PDM. 3. Open the battery door compartment and remove the battery (if present), for approximately 10 seconds. 4. Reinsert the battery
Quantity111,728 units
Distributionworldwide distribution US nationwide, Albania, Algeria, Argentina, Australia, Austria, Bahamas, Bahrain, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Fij
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071073SOLAR 8000 AND TRANSPORT PRO WITH PATIENT DATA MODULE
Ge Medical Systems Information Technologies, May 11, 2007