ClearanceBrief
Recall Z-2152-2021 ยท LPL Lenses, Soft Contact, Daily Wear

CooperVision Inc.

CLARITI 1 DAY TORIC Power -07.00 -1.25/010 Daily Disposable Contact Lens with UV blocker

DateMay 19, 2021
StatusTerminated
ReasonMisaligned placement of an axis mark can cause the patient to experience blurred vision
Root causeEquipment maintenance
ActionCoopervision issued Urgent Medical Device Correction letter on 5/19/21 to wholesale/distributor and Eye Care Practitioner level. The independent accounts via overnight FedEx (receipt tracking). Distributors and corporate accounts via email, with a read receipt. Letter states reason for recall, health risk and action to take: Examine your inventory for the affected lot numbers on the enclosed Ship Report. " Stop any further distribution of the specified lots of clariti 1 day toric lenses. " Dispose of the product and, within the next five (5) business days, please complete the attached Corre
Quantity450 lenses
DistributionWorldwide distribution - US Nationwide distribution in the states of FL, NY, TN, OH, NC, WI and the countries of France, United Arab Emirates, Brazil, Australia, New Zealand, Taiwan, Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K130331SAUFLON CLARITI 1 DAY (SOMOFILCON A) SOFT (HYDROPHILIC) DAILY DISPOSABLE CONTACT LENS WITH UV BLOCKER
Sauflon Pharmaceuticals, Ltd., Jul 26, 2013