Recall Z-2076-2021 ยท MXD Recorder, Event, Implantable Cardiac, (With Arrhythmia Detection)

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Medtronic LINQ II Model Number # LNQ22 Insertable Cardiac Monitor Sterile EO Rx Only

DateJul 14, 2021 (initiated May 14, 2021)
StatusOpen, Classified
ReasonDevice fail to detect Brady episodes, Pause (Asystole) episodes and PVC's after a ride-through POR reset.
Root causeSoftware design
ActionOn June 1 2021 Medtronic sent a letter notifying the consignees the following: -LINQ II insertable cardiac monitors (ICM) that undergo a partial electrical reset appear to be programmed ON, but are no longer able to detect and report Brady, Pause, and PVC events to clinicians. Medtronic estimates that 0.21% of LINQ II ICMs have experienced a partial electrical reset resulting in the inability to detect Brady, Pause, and PVC events. -A correction for currently implanted LINQ II ICMs is not available. -We are requesting that hospitals quarantine all LINQ II ICMs on hospital shelves. Physicians
Quantity21,685
DistributionWorldwide distribution - US Nationwide in the states of AK, AL, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY. The countries of Australia, Austria, Canada, Denmark, Finland,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K200795LINQ II Insertable Cardiac Monitor, LINQ Mobile Manager, Device Command Library, Instrument Command Library, LINQ Tool K
Medtronic, Inc., Jul 3, 2020