ClearanceBrief
Recall Z-1958-2021 ยท MNS Ventilator, Continuous, Non-Life-Supporting

Philips Respironics, Inc.

A-Series BiPAP A 40, A-Series BiPAP A30 Continuous Ventilator, Non-life Supporting

DateJul 15, 2021 (initiated Jun 14, 2021)
StatusOpen, Classified
ReasonThe polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the devices air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.
Root causeDevice Design
ActionUrgent Medical Device Recall letters were sent to customers starting 6/14/21. 1. Do not stop or alter your prescribed therapy until you have talked to your physician. Your physician may evaluate if the benefit outweighs the risks. 2. If your physician determines that you must continue using this device, use an inline bacterial filter. Your Instructions for Use has guidance on installation. 3. Register your device on www.philips.com/src-update or call 1-877-907-7508. An updated URGENT: Medical Device Recall notification letter was issued on 11/16/21. 1. Do not stop or change ventilator use
Quantity72665
DistributionWorldwide distribution including US Nationwide, Argentina, Australia, Brazil, Canada, China, France, Germany, Italy, Japan, Korea, and Latin America.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113053BIPAP A 30 VENTILATORY SUPPORT SYSTEM
Respironics, Inc., Feb 1, 2012
K121623BIPAP A 40 VENTILATORY SUPPORT SYSTEM
Respironics, Inc., Dec 14, 2012