Recall Z-1974-2021 ยท BZD Ventilator, Non-Continuous (Respirator)

Philips Respironics, Inc.

SystemOne (Q-Series), DreamStation, DreamStation Go, Dorma 400, Dorma 500, REMstar SE Auto Noncontinuous Ventilators The Philips Respironics E30 ventilator is intended to provide invasive and non-invasive ventilatory support for individuals with Respiratory Insufficiency. It is specifically for the care of adult and pediatric patients >7 years of age and >18kgs.

DateJul 13, 2021 (initiated Jun 14, 2021)
StatusOpen, Classified
ReasonThe polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the device s the air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.
Root causeDevice Design
ActionUrgent Medical Device Recall notification letters were sent to customers beginning on 6/14/21. Immediate Actions to be taken by You, the User: 1. Discontinue use of your device and work with your physician or Durable Medical Equipment (DME) provider to determine the most appropriate options for continued treatment. 2. Register your device on the recall website www.philips.com/src-updates Call 1-877-907-7508 if you cannot visit the website or do not have internet access. An updated URGENT: Medical Device Recall notification letter was issued on 11/16/21. Immediate Actions to be taken by Y
Quantity15,357,775 (10,307,186 in US, 5,039,748 OUS) in total
DistributionGlobal distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K091319REMSTAR AUTO A-FLEX
Respironics Inc., Sleep & Home Respiratory Group, Aug 3, 2009
K113068REMSTAR AUTO A-FLEX HT
Respironics, Inc., Dec 16, 2011
K122769REMSTAR SE
Respironics, Inc., Oct 9, 2012
K130077REMSTAR SE
Respironics, Inc., May 21, 2013
K131882GEODESICEEG SYSTEM 400 SERIES (GES 400)
Electrical Geodesics, Inc., Feb 12, 2014
K131982REMSTAR AUTO A-FLEX HT
Respironics, Inc., Oct 18, 2013