Recall Z-1994-2021 ยท LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days

Bard Access Systems Inc.

5 Fr IR PowerPICC Convenience Kit Catheter Dual Lumen with 70 cm Guidewire Unit Label: PowerPICC REF CK000340B BARD MYPICC KIT 5F French Size Dual-Lumen 70cm Guidewire

DateMay 19, 2021
StatusTerminated
ReasonCatheter convenience kits contain the incorrect guidewire. Kits were labeled with 70cm guidewires, but contained a 50cm guidewire.
Root causeProcess control
ActionOn 05/19/2021, the Recalling Firm sent an "URGENT MEDICAL DEVICE CORRECTION" Notification via FedEx 2-day to customers informing them that the Recalling Firm has confirmed through customer complaint that a number of catheter convenience kits labeled as having 70cm guidewires may actually contain 50cm guidewires. The Recalling Firm is providing customers with a quick way of identifying if their kit contains the correct or incorrect size guidewire by determining the number of loops of the wire. Customers are instructed to: 1. Immediately review their inventory for the specific Catalog (Ref) an
Quantity30 kits
DistributionU.S. Nationwide distribution in the state of RI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.