ClearanceBrief
Recall Z-2033-2021 ยท PUI Drape, Surgical, Exempt

Cardinal Health 200, LLC

Sterile Radial/Femoral Angiography Drape, Catalog Number 29529 - Product Usage: intended to be used as protective patient coverings, such as to isolate the site of a surgical incision from microbial and other contamination.

DateMay 21, 2021
StatusTerminated
ReasonThe plastic film used to manufacture these Angiography Drape and Cesarean Birth Drape products is causing the drapes to stick, which may prevent the products from being unfolded and applied properly. Sold in single sterile and bulk non-sterile configurations.
Root causeNonconforming Material/Component
ActionThe firm initiated the recall by letter on 05/20/2021. The letter explained the issue and the hazard, and requested the return of the units.
Quantity3192 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.