Recall Z-1976-2021 ยท MIH System, Endovascular Graft, Aortic Aneurysm Treatment

Medtronic Vascular, Inc.

Endurant II/IIs Stent Graft System: STENT GRAFT ETBF3216C145E ENDUR II BIF. For treatment of vascular aneurysms.

DateApr 28, 2021
StatusTerminated
ReasonIncorrect stentstop assembly (18fr) was used to build the delivery system component of the device. Correct stentstop assembly is 20fr.
Root causeProcess control
ActionBeginning on 04/28/2021, the Recalling Firm initiated a verbal notification and retrieved affected product for a single U.S. customer. On 05/10/2021, the Recalling Firm sent an "URGENT MEDICAL DEVICE RECALL" Notification Letter via UPS 2-day delivery to inform its impacted customer that the Recalling Firm has determined that the stent graft system was manufactured with an incorrect stentstop assembly. The letter was also to to ask for confirmation that the customer no longer had the affected devices in its possession. The letter instructed customers to take the following actions: 1. Complet
Quantity3 devices
DistributionU.S. Nationwide distribution in the state of GA. O.U.S.: None
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.