Recall Z-1897-2021 ยท QAN Stent, Iliac Vein

Bard Peripheral Vascular Inc

VENOVO Venous Stent System 10F Product Description(REF)/dimension: VENEL16040 Venovo 10F 16/40/1200mm OUS VENEL16060 Venovo 10F 16/60/1200mm OUS VENEL16080 Venovo 10F 16/80/1200mm OUS VENEL16100 Venovo 10F 16/100/1200mm OUS VENEL16120 Venovo 10F 16/120/1200mm OUS VENEL16140 Venovo 10F 16/140/1200mm OUS VENEL16160 Venovo 10F 16/160/1200mm OUS VENEL18040 Venovo 10F 18/40/1200mm OUS VENEL18060 Venov

DateDec 14, 2020
StatusOpen, Classified
ReasonReports of the proximal end of the stent not immediately expanding upon deployment and remaining connected to the delivery system.
Root causeUnder Investigation by firm
ActionOn 01/13/2021, the firm sent an "URGENT MEDICAL DEVICE SAFETY NOTICE" via FedEx to customers informing them that the Recalling Firm has received reports regarding the venous stent system 9F where the proximal end of the stent does not immediately expand upon deployment, and remains connected to the delivery system, and that over-manipulation or forcing the catheter delivery system in attempts to assist the stent's expansion, could potentially have a varying degree of harm associated with it. potential harm ranges from prolonging the procedure, damaging or deformity of the stent, potential vas
QuantityTotal Devices=57,515 (U.S.=46,878; O.U.S.=10,637)
DistributionWorldwide distribution. US Nationwide, ALGERIA, ANDORRA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELGIUM, BRAZIL, BRUNEI, BULGARIA, CHILE, COSTA RICA, CYPRUS, CZECH REPUBLIC, DENMARK, DEUTSCHLAND, EGYPT, FINLAND, FRANCE, GEORGIA, GREECE, HONG KONG, HUNGARY, INDIA, IRAN, IRAQ, IRELAND, I
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.