ClearanceBrief
Recall Z-1796-2021 ยท LNH System, Nuclear Magnetic Resonance Imaging

Philips North America Llc

Ingenia Ambition S - Model Number: 781359 Diagnostic MR system.

DateJun 4, 2021 (initiated Apr 23, 2021)
StatusOpen, Classified
ReasonThe MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. during decommissioning or moving activities. A mechanical issue was detected on these hoisting interface blocks on top of the magnet.
Root causeDevice Design
ActionOn March 23, 2021, Philips issued an Urgent Medical Device Correction notice to customers via letter stating Philips identified an issue with specified MR systems. The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. during decommissioning or moving activities A mechanical issue was detected on these hoisting interface blocks on top of the magnet. In the initial design of the hoisting blocks, the diameter to meet the required bolt-on interface surface for the swivel
Quantity156 MR systems
DistributionDomestic: FL, GA, MI, MO, NH, NJ, NY, PA, SC, TX, VT. Foreign: Australia, Austria, Bangladesh, Belgium, Brazil, Denmark, Finland, France, Germany, Iraq, Italy, Japan, Libya, Moldova, Netherlands, Oman, Spain, Switzerland, Thailand, United Arab Emirates, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181479Ingenia Ambition S and Ingenia Ambition X
Philips Medical Systems Nederland B.V., Aug 3, 2018