Recall Z-1862-2021 ยท DTN Reservoir, Blood, Cardiopulmonary Bypass

Circulatory Technology Inc.

The Better-Bladder (BB), part of extracorporeal membrane oxygenation (ECMO) cardiopulmonary bypass circuits. Model: BB14

DateApr 27, 2021
StatusOpen, Classified
ReasonThe device is used as part of extracorporeal membrane oxygenation (ECMO) cardiopulmonary bypass circuits and may collapse during use which can increase resistance to flow in the venous line and cause a drop in circuit blood flow.
Root causeProcess control
ActionOn April 27, 2021, CTI issued an Urgent Medical Device Removal notice to customers via email and phone, notifying them three better bladder BB14 units were found to have the pigtails dislodge from the housing. We are requesting that you immediately inspect your inventory, and quarantine all units having the aforementioned lots. These units must be returned to CTI. CTI will provide UPS shipping labels for returning the units and will provide replacements or credit for the returned units. Replacement BB14 from an unaffected lot will be available for shipping early next week. However, if you n
Quantity147 units
DistributionUS Nationwide distribution in the states of AL, AR, CA, DC, FL, GA, IN, MO, NC, PA, SC, TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K981284BETTER-BLADDER (BB)
Circulatory Technology, Inc., Aug 12, 1998