Recall Z-1968-2021 ยท FAP Cystometric Gas (Carbon-Dioxide) On Hydraulic Device

Spes Medica

SNS Disposable Urethral Catheter Electrode, 10 Fr, Gold Recording/Stimulating Contact printed on polyurethane film adhesive with attached 2 meter lead-wire with touch-proof connector, Reference Number UE006

DateApr 8, 2021
StatusTerminated
ReasonThe supplier informed them of possible deviations of the parameters/processes defined for ethylene oxide sterilization.
Root causeMaterial/Component Contamination
ActionAn email titled "Object: Spes Medica Safety Notice for possible Non-Conformities to Steril Milano" dated 4/8/21 was sent to customers. Following the communications of Steril Milano and the Ministry of Health (http:ljwww.salute.gov.it/portale/news/p3 2 1 1 1.jsp ?lingua=italiano&menu=notizie&p=dalministero&id=5365) on possible deviations of the parameters/processes defined for ethylene oxide sterilization, already validated and routinely used by Steril Milano; the company Spes Medica S.r.I. informs you that in the years 2019-2020-2021 has used the Steril Milano supplier only once in- August
Quantity34 boxes
DistributionDistribution in the following US states: CA, MA, and SC.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113771SIGNAL GEAR URETHRAL CATHETER ELECTRODE
Neurovision Medical Products, Inc., Dec 13, 2012