Recall Z-2082-2021 ยท KPO Dialysate Concentrate For Hemodialysis (Liquid Or Powder)

Fresenius Medical Care Holdings, Inc.

(1) 08-4078-BB BIBAG 650 GRAMS/BAG 16 BAGS/CASE (2) 08-4080-BB BIBAG 900 grams/bag 12 bags/case For patients with acute and chronic end stage renal disease going under PD in healthcare or home setting GRANUFLO & NATURALYTE: Concentrate is formulated to be used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kid

DateApr 28, 2021
StatusTerminated
ReasonExposed to freezing temperatures due to refrigerated truck malfunctions
Root causeEnvironmental control
ActionFresenius issued telephone script communication on April 28.2021 stating reason for recall, heath risk and action to take: Tell them you already contacted their Home Therapy Nurse to alert them that you would be calling them. o The purpose of my call today is that you received a delivery from us on (either Feb 22nd or Feb 23rd). o The truck you received a delivery from had a temperature malfunction due to the bad storms Texas experienced back in February. This caused the product you were delivered to be exposed to freezing or near freezing temperatures for about 3 hours. o o I see you receive
Quantity(1) 20 cases (2) 30 cases
DistributionUS Distribution to: TX only
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K162716bibag
Fresenius Medical Care Renal Therapies Group, LLC, May 18, 2017